Electrochemical Detector SHE-8

Electrochemical Detector

In the case of applied voltage, the electrochemical detector detects the change of current caused by the redox reaction of the substance to be measured on the electrode surface. Electrochemical detector is often used to analyse ions with low dissociation, which are difficult to detect with conductivity detectors and have electrical activity .

Technology Advantage
Unique noise reduction technology
Advanced Digital Filter (ADF), different from traditional filter time constant (FTC), has less signal loss in the process of noise reduction, which can let low frequency peaks pass, remove high frequency noise, improve signal-to-noise ratio and increase detection limit.

Unique structure of analytical flow cell

Central spray wall structure. The sensitivity of analytical flow cell can be continuously adjusted. The working volume can be continuously adjusted between 0 and 300 nl.One electrode can meet the application requirements of different sensitivity.

Built-in Faraday shielded temperature oven

Reduce static and unexplained interference and influence of temperature change, improve repeatability and stability.

Applicable to IC, HPLC and UHPLC
With DC ,PULSE and SCAN mode
The most sensitive electrochemical detector
Different flow cells to cover every application

Mode
DC ,PULSE and SCAN

Working Temperature
10-40℃(indoor use only)

Safety and EMC
According to Ec-directives; Emission Group I Class A;cMETus approved

Sensors
Up to 3 flow cells

Max Current Compensation(Autozero)
– 25 nA-25 mA in DC and PULSE mode dependent on range setting

Oven
+7°C above ambient to 60°C, accuracy 0.5°C, stability 0.1°C; accommodates column and flow cell(s)

Offset
+50% to – 50% of max. output voltage, 5% steps

Analog Output (DAC)
-1 to +1 V full scale (via 16-bit D/A converter)

Analog Output (I/E)
-1 to +1 V full scale (via 16-bit D/A converter)

DC mode

Range
10 pA-200 uA in 1. 2. 5 increments

Filter (ADF)
10-0.001 Hz in 1.2. 5 increments
RAW and OFF: for unprocessed data

Potential ( EC)
-2.50V to +2.50V with 10 mV increments

Data Rate
1- 100 Hz in 1, 2, 5 increments, dependent on filter setting

Noise
<2pA with dummy cell(load of 300 MΩ /470 pF) in 1 nA range, filter off, Ec+800mV and temperature of 35’C

PULSE mode

Range
10 nA – 200 μA in 1, 2, 5 increments
OFF: for unprocessed data

Filter (ADF)
0.5 – 0.001 Hz in 1, 2, 5 increments

Potential ( EC)
-2.50 V to + 2.50 V with 10 mV increments

Data Rate
1/(pulse duration) Hz

SCAN mode

Range
10 nA – 200 μA in 1, 2, 5 increments
OFF: for unprocessed data

Potential (Ec)
-2.50 V to + 2.50V with 10 mV increments

Data Rate
1 Hz Scan rate 1 – 100 mV/s in 1, 2, 5 increments

Others

Cycle
Half, Full, Continuous

Dimensions
-43(D)*22(W)*44(H)cm

Power requirements
100-240 VAC, 50/60 Hz, 260 VA, auto-sensing

Weight
Max 14. 4 kg without flow cell and column